CareFusion Alaris 8015 User Manual page 411

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Alarm Message
HIGH ETCO2
HIGH FICO2
HIGH RR
HIGH SATURATION
HIGH_PULSE
INCOMPATIBLE
MODULE
INFUSION
COMPLETE
INFUSION
COMPLETE - KVO
INSTALL PRESSURE
DISC
LOW BATTERY
< 30 MIN
PLUG IN NOW
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Alarms and Alerts (continued)
Alarms (continued)
Device
Priority
EtCO
High
2
EtCO
High
2
EtCO
High
2
SpO
High
2
SpO
High
2
PC unit
Informational
Syringe
Medium
PCA
Pump
Pump
Medium
Syringe
Syringe
High
PC unit
Low
Alarms
Meaning
The EtCO
value is above the
2
specified limit.
FiCO
value is above the
2
specified limit.
The respiratory rate is above the
specified limit.
The high saturation alarm limits
have been exceeded.
The high pulse rate alarm limit
has been exceeded.
The PC unit is unable to
recognize the attached
MODULE because it is an
unsupported module type
The current infusion is
completed with no KVO.
VTBI has been infused; the
module is infusing at the KVO
rate
The Pressure Sensing Disc is
removed. If an infusion is
running, the infusion stops on
affected module.
Low battery threshold sensed;
the remaining battery run time is
limited.
Response
Assess the patient's condition. Confirm
that the correct alarm limit values are
selected.
Assess the patient's condition. Confirm
that the correct alarm limit values are
selected.
Assess patient's the condition. Confirm
that the correct alarm limit values are
selected.
Assess the patient's condition. Confirm
that the correct alarm limit values are
selected.
Assess the patient's condition. Confirm
that the correct alarm limit values are
selected.
Remove the unsupported module.
Set up a new infusion or press the
CHANNEL OFF key.
Set up a new infusion, if required, and
reinititate the infusion
Reinstall the Pressure Sensing Disc
and press the RESTART key
Connect to a power source (alarm
silenced). Low battery will continue to
display after the AC is plugged in until
the battery has built up sufficient
charge to run the Alaris System for 30
minutes. To verify AC is charging, look
at the AC LED on the front panel and
verify that is is on.