CareFusion Alaris 8015 User Manual page 85

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Prepare Syringe and Administration Set (Syringe Module) (Continued)
Prime—Using Options Menu (Continued)
Administration Set With Pressure Sensing
Disc (Continued)
c. Apply firm upward pressure on pressure sensing disc
(not tubing) until disc snaps into place.
12. To return to main screen, press
If
soft key is pressed before pressure sensing disc is
EXIT
reinstalled, a prompt to reinstall pressure sensing disc is
displayed.
Administration Set With No Pressure
Sensing Disc
1.
Press
key.
OPTIONS
2.
Press Prime Set with Syringe soft key.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
soft key.
EXIT
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pressure Sensing Disc
A A A
Re-install Pressure Sensing
Disc.
>Press CONFIRM
CONFIRM
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - 550 mmHg
Prime Set with Syringe
>Select an Option or EXIT
PAGE
EXIT
DOWN
Getting Started
2-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents