Manual Programming—Primary Infusion (Continued)
Fluid Infusion (Continued)
Rate/Volume Infusion (Continued)
3.
Verify correct infusion parameter entry and press
soft key.
If the programmed
that care area, an audio alert sounds and a visual
prompt appears before programming can continue. If
Yes soft key is pressed, programming continues; if No
soft key is pressed, infusion needs to be
reprogrammed.
If the programmed
that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
Infusion needs to be reprogrammed.
If a Soft Limit is overridden, G icon is displayed. When
G soft key is pressed, all applicable out-of-range limits
are listed.
4.
Syringe module:
Unclamp tubing and attach administration set to
patient.
FU
RATE (mL/h)
SELECT
CHANNEL
Unclamping tubing and starting infusion before
PAUSE
CHANNEL
OFF
RESTART
attaching administration set to patient minimizes
any potential bolus that can be released from
pressure built up in set due to normal syringe
loading and priming.
Volume/Duration Infusion
1.
Press
VOLUME DURATION
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
fluid is outside the Soft Limit for
IV
fluid is outside the Hard Limit for
IV
soft key.
START
Manual Programming with Guardrails Suite MX
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Guardrails Fluid Setup
TPN
PRIMARY INFUSION
_ _ _
RATE
mL/h
_ _ _
mL
VTBI
>Select RATE
VOLUME
DURATION
2-47