Programming with Interoperability—Secondary Infusion (Pump Module)
Stop Secondary and Return to Primary
1.
Press the
CHANNEL SELECT
2.
Press the
SETUP
3.
Press the
PRIMARY
4.
Close the clamp on the secondary administration set.
-or-
Disconnect the secondary administration set from the upper
injection port.
5.
Press the
START
6.
To stop the secondary infusion and begin infusing the
primary, press the
The secondary infusion stops and primary infusion begins.
The main screen appears.
NOTE:
The SEC to PRI alert does not sound when the infusion is
manually ended and returned to the primary.
Programming with Interoperability and Guardrails Suite MX
2-86
(Continued)
key.
soft key.
soft key.
soft key.
soft key.
Yes
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015