Programming with Interoperability and Guardrails Suite MX

Interoperability is designed to help automate existing manual
workflows with regard to programming infusions on the Alaris
Pump and Syringe modules.
Interoperability refers to the ability of the Alaris System Pump
module and Syringe module to:
Receive infusion order parameters from a third-party
system for pre-population. This may be referred to as an
automated programming request.
Publish infusion status messages for consumption by
third-party systems.
The following information on programming an infusion with
interoperability does not include information regarding
publishing infusion status messages from the Alaris System.
For information on infusion status messages, refer to the
following four guides:
Infusion Status Message Specification for CareFusion
Coordination Engine—Regulated Infusion Adapter v1.3
HL7 Order Message Specification for CareFusion
Coordination Engine—Regulated Infusion Adapter v1.3
Alaris System Interoperability Services User Reference
Guide
Alaris System Interoperability Services Guide for Clinical
Workflow Testing
The Alaris System with interoperability enables pre-population
of infusion parameters from an infusion order in the hospital's
electronic medical record/hospital information system
(EMR/HIS) using the EMR/HIS vendor barcode medication
administration (BCMA) system.
Pre-population of infusion parameters reduces the number of
programming screens and key presses required with manual
programming. There are no differences in programming
screens, prompts or the user interface between an Alaris
System with interoperability and an Alaris System without
interoperability. The implementation of interoperability does not
preclude a clinician from manually programming the Alaris
System. Manual programming is required in the event of a
failure in any component of the interfaced system.