Guardrails Software; Administration Set/Syringe Information - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

WARNING
The Guardrails Suite MX incorporates dosing limits
and instrument configuration parameters based on
hospital/facility protocol. The software adds a test of
reasonableness to drug programming based on the
limits defined by the hospital/facility. Qualified personnel
must ensure the appropriateness of drug dosing limits,
drug compatibility, and instrument performance, as part
of the overall infusion. Potential hazards include drug
interactions, inaccurate delivery rates and pressure alarms,
and nuisance alarms.
When loading a Data Set with the Guardrails Suite MX,
ensure that the correct Profile (for patient care area) is
selected prior to starting an infusion. Failure to use the
appropriate Profile could cause serious consequences.

Administration Set/Syringe Information

The PCA module uses standard, single-use, disposable
syringes (with luer-lock connectors) and administration sets
with anti-siphon valves, designed for use on syringe-type PCA
pumps.
For specific administration set instructions and set
replacement interval, refer to User Manual provided with set.
For a list of compatible syringes, refer to Compatible
Syringes on page 3-49.
Use aseptic techniques when handling sets and syringes.
Administration sets are supplied with a sterile and
nonpyrogenic fluid path for one-time use. Do not
resterilize.
Discard administration set per facility protocol.
For
push medication (put instrument on hold), clamp
IV
tubing above port.
Flush port(s) per facility protocol.
General Information
3-48
Warnings and Cautions (Continued)

Guardrails Software

Alaris System User Manual – with v9.33 Model 8015
PCA Module Section

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents