Trumpet and Start-Up Curves (Continued)
Pump Module (Continued)
Under conditions of +300 mmHg pressure, the Pump module
typically exhibits a long-term accuracy offset of approximately
-6.2% from mean values.
Under conditions of -100 mmHg pressure, the Pump module
typically exhibits a long-term accuracy offset of approximately
+0.5% from mean values.
Resulting trumpet observation points typically track those
of accuracy; therefore, no significant change in short-term
variations result under these pressure conditions.
Effects of Negative Solution Container Heights
With a negative head height of -0.5 meters, the Pump module
typically exhibits a long-term accuracy offset of approximately
-2.8% from mean values.
Resulting trumpet observation points typically track those
of accuracy; therefore, no significant change in short-term
variations result under negative head height conditions.
Trumpet and start-up curves have been provided for 1.0 mL/h
and 5.0 mL/h. Measurements for trumpet curve rates below
1.0 mL/h are not provided because of the difficulty in
measuring extremely small volumes over a large duration of
time. In this case, the linear relationship of the plunger position
and velocity to syringe volume and rate is verified, and is a
function of the accuracy of the design.
Measurements for trumpet curve rates above 5.0 mL/h are also
not provided, as the syringe volume is displaced in a very short
time with a rate up to 999 mL/h. Accuracy, however, is assured
with the design implementation.
Under conditions of -100 mmHg, +100 mmHg, and +300 mmHg
pressures, the Syringe module typically exhibits a long-term
accuracy offset of approximately 0.2% or less from the mean
value.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015

Syringe Module