Accumulated Air Window:
Bolus Volume, Maximum
after Occlusion:
Critical Volume:
Dimensions:
Environmental Conditions:
Equipment Orientation:
General Information
2-152
Specifications
Pump Module
Single Bolus
Setting (mcL)
50
75
250
500
NOTE:
In a standard bore tubing set:
50 mcL is approximately 0.34 inches of air
75 mcL is approximately 0.51 inches of air
250 mcL is approximately 1.7 inches of air
In Anesthesia Mode only.
Pressure Limit (mmHg
50
525
Maximum over-infusion that can occur in the event of a single fault condition is
0.6 mL.
3.3" W x 8.9" H x 5.5" D
Atmospheric Pressure
Relative Humidity
(Avoid prolonged
exposure to relative
humidity >85%)
Temperature Range
To ensure proper operation, Alaris System must remain in an upright position.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Volume Window
% Air that Causes
(mL)
2.8
8.0
8.0
12.0
Rate (mL/h)
25
25
Operating
525 - 4560 mmHg
(700 - 6080 hPa)
20 - 90%
Noncondensing
41 - 104°F
(5 - 40°C)
Alaris System User Manual – with v9.33 Model 8015
Volume (mL) that
Alarm
Causes Alarm
10%
20%
30%
30%
Bolus Volume (mL)
0.3
0.6
Storage/Transport
375 - 760 mmHg
(500 - 1013 hPa)
5 - 85%
Noncondensing
-4 - 140°F
(-20 - 60°C)
Pump and Syringe Modules Section
0.28
1.6
2.4
2.4