Programming with Interoperability—Initial Primary Infusion (Continued)
Intermittent Infusion (Continued)
3.
Review and verify that all parameters pre-populated on the
Alaris System are correct prior to starting the infusion.
The rate, VTBI, and duration are pre-populated from the
automated programming request.
4.
Press the
START
5.
For the Syringe module, unclamp tubing and attach
administration set to patient.
NOTE:
Unclamping tubing and starting infusion before attaching
administration set to patient minimizes any potential bolus
that can be released from pressure built up in set due to
normal syringe loading and priming.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
soft key.
Programming with Interoperability and Guardrails Suite MX
A A A
Guardrails Drug Setup
Potassium Chloride
PRIMARY INFUSION
50
RATE=
mL/h
50
VTBI
mL
1:00
DURATION
hh:mm
0.143
DOSE =
mEq/kg
0.2 mEq/kg
[Conc]:
>Press START
DELAY
RATE
SETUP
START
OPTIONS
VOLUME