Respiration Rate Test
The following test method is used to determine the module's
rated respiration rate range and the corresponding effects of
end-tidal gas readings accuracy as a function of respiratory
rate:
•
The system is set for operation.
•
The power connection is verified.
•
The module is connected to the PC and power supply.
•
A scope is connected to the digital/analog multifunction I/O.
•
A digital output scheduler software is opened on the PC.
•
Gas connections are checked.
•
The power connection is verified.
•
The gas mixes cylinders are open, set on the required
pressure, and their condition is verified.
•
A BPM setting is selected.
•
The module is connected to the gas tube.
•
The module is disconnected.
NOTE:
When CO
waveforms are generated the CO
2
increases in height and then returns to the selected baseline rate
(BPM).
EtCO
Module Section
2
Alaris System User Manual – with v9.33 Model 8015
Measurement Accuracy (continued)
reading line
2
General Information
5-25