Advertisement

Features and Definitions (Continued)
Feature
NEOI Snooze
Numeric Pressure
Display
Occlusion Pressure
Pressure Sensing Disc
Priming
Syringe Empty
Syringe Volume
Detection
General Information
2-142
Features and Displays (Continued)
Optional capability to remind users NEOI alarm has been silenced. This is
shared Syringe and PCA setting.
Dynamic pressure display is numerical and only available when pressure
sensing disc is inserted.
A complete range of downstream occlusion detection options is provided.
• With pressure sensing disc: Downstream occlusion alarm threshold is
selectable between 25 and 1000 mmHg, in 1 mmHg increments.
• Without pressure sensing disc: Downstream occlusion alarm threshold
can be set to low, medium, or high.
When installed, pressure sensing disc significantly improves instrument's
pressure sensing capabilities for a faster occlusion detection time, and
makes following features available:
Auto Pressure
Back-Off
Customizable Pressure Alarm Settings (see "Occlusion Pressure")
Fast Start
Pressure Tracking
Allows a limited volume of fluid to be delivered in order to prime
administration set prior to being connected to a patient or after changing
a syringe. When priming, a single continuous press of
delivers up to 2 mL of priming fluid.
Instrument gives an alert and stops when an empty syringe is detected.
System automatically detects fluid volume in a syringe when it is inserted.
Definition—Syringe Module
Alaris System User Manual – with v9.33 Model 8015
soft key
PRIME
Pump and Syringe Modules Section
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART