Volume/Duration Disabled (Continued)
6.
Syringe module:
•
Unclamp tubing and attach administration set to
patient.
FU
RATE (mL/h)
SELECT
CHANNEL
•
Unclamping tubing and starting infusion before
PAUSE
CHANNEL
OFF
RESTART
attaching administration set to patient minimizes
any potential bolus that can be released from
pressure built up in set due to normal syringe
loading and priming.
7.
To see detail screen during or between infusions, press
CHANNEL SELECT
•
During infusion, Volume Remaining is displayed.
•
Between infusions:
◦
Number of doses completed and when next dose
starts are displayed.
◦
Yellow Standby Status Indicator illuminates.
NOTES:
See "Delay Options" procedure to program an infusion delay.
When delaying an infusion, a multidose cannot be delayed for
more than 8 hours, and all doses in the multidose program
must be completed within a 24-hour program.
Syringe module: If NEOI is enabled, the Near End of infusion
message appears near the end of the last dose.
General Programming
2-112
Multidose Mode (Continued)
key.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART