Prepare Administration Set (Pump Module) (Continued)
c. Press safety clamp fitment into recess below
mechanism.
d. Using a finger tip, firmly push tubing toward back of
Air-in-Line (AIL) Detector.
5.
Close door and latch, as follows:
a. Close door and hold in a closed
position by grasping door and
instrument case with one hand.
b. Gently lower latch.
Safety clamp device is automatically
disengaged.
6.
Open roller clamp.
7.
Verify that no fluid is flowing through drip chamber.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Load (Continued)
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
CAUTION
To reduce the potential for nuisance
alarms, ensure that tubing is
AIL
fully inserted in the AIL Detector.
Safety Clamp Fitment
Push tubing toward
back of AIL Detector.
WARNING
• Do not touch the administration
set while closing the door. Failure
to follow this instruction can result
in infusion rate inaccuracy.
• To prevent a potential free-
flow condition, ensure that no
extraneous object (for example,
bedding, tubing, glove) is
enclosed or caught in the Pump
module door.
Getting Started
2-5