Electromagnetic Environment (Continued)
Alaris System (Continued)
Immunity Test
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Compliance (Continued)
Electromagnetic Immunity - Life Support Equipment
IEC 60601-1-2
Test Level
10 Vrms
10 Vrms
150 kHz -
80 MHz
10 V/m
10 V/m
80 MHz -
2.5 GHz
Table 3
Compliance
Level
Portable and mobile RF communications
equipment should be used no closer to
Alaris System (including cables) than
recommended separation distance calculated
from equation applicable to frequency of
transmitter.
Recommended Separation Distance:
__ d = [-----] √P
__ d = [-----] √P 80 MHz - 800 MHz
__ d = [-----] √P 80 MHz - 2.5 GHz
P = maximum output power rating of transmitter
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than compliance level in each
Interference might occur in vicinity of
equipment marked with following symbol:
Electromagnetic Environment -
Guidance
12
V
2
12
E
1
12
E
1
d = recommended separation distance in
meters (m).
in watts (W) according to transmitter
manufacturer.
frequency range.