CareFusion Alaris 8015 User Manual page 413

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Alarm Message
NEAR END OF
INFUSION
NEOI Snooze
NETWORK
COMMUNICATION
ERROR
NO BREATH
DETECTED
NO SENSOR
NO SENSOR
NO SIGNAL
OCCLUDED - FLUID
SIDE/EMPTY
CONTAINER
OCCLUDED -
PATIENT SIDE
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Alarms and Alerts (continued)
Alarms (continued)
Device
Priority
Syringe,
Low
PCA
Syringe
Reminder
PCA
PC unit
Informational
EtCO
High
2
SpO
Informational
2
SpO
Medium
2
SpO
Medium
2
Pump
High
Pump
High
Alarms
Meaning
The syringe is almost empty.
This is a timed event that can be
configured in the Data Set. To
set or change this option, see
"General Information" and
"Configurable Settings".
A periodic audible signal that
reminds the user that a Near
End of Infusion alarm has been
previously silenced.
Wireless connections is lost.
No breath has been detected for
a specified period of time.
No sensor is connected to the
module while attempting to
search for a patient.
No sensor is connected to the
module during monitoring or
after approximately 30 seconds
of searching.
Failure to find a patient signal
Indicates either an upstream
occlusion or an empty container.
The infusion stops on affected
module.
Increased back pressure sensed
while infusing. The infusion
stops on affected module.
Response
To silence the alarm, press the
SILENCE key. The module remains
functional and continues the infusion.
None
The PC unit continues to operate
without wireless functionality.
Assess the patient's condition. Confirm
that the correct alarm limit values are
selected. Check the Microstream
disposable to confirm that the correct
disposable is chosen. Confirm the
correct disposable placement.
Attach the sensor to the module.
Check the patient. Attach the sensor to
patient.
Check the sensor. Confirm the correct
sensor placement.
Clear the occlusion on fluid side of the
device. If necessary, refill the drip
chamber. Press the RESTART key, or
press the CHANNEL SELECT key and
then the START soft key.
Clear the occlusion. Press the
RESTART key, or press the
CHANNEL SELECT key and then the
START soft key.
A-19