Display references throughout this procedure are for illustration
purposes only.
See "General Information," "Features and Displays" and the
PC unit section of this User Manual for information about:
Displays
Operating Features, Controls, Indicators
The majority of user interface programming is identical for both
SpO
modules.
2
1.
Perform following steps (see PC unit section of this User
Manual, "General Setup and Operation," "Start-Up"):
a. Power on system.
b. Choose Yes or No to New Patient?
c. Confirm current Profile or select a new Profile.
d. Enter patient identifier, if required.
2.
Attach patient cable and sensor (see "Getting Started").
3.
Press
CHANNEL SELECT
SEARCHING
display until SpO
(approximately 15 seconds).
If sensor is not attached to a site,
displayed.
To prevent screen from reverting to Main Display,
press
ENTER
screen is displayed.
If sensor is not attached during message display,
module goes into sleep mode. To begin monitoring
once module is in this mode, press
4.
Ensure that sensor's red LED is on.
SpO
Module Section
2
Alaris System User Manual – with v9.33 Model 8015

Monitoring Mode

key.
might appear in Channel Message
and pulse readings stabilize
2
SENSOR OFF
key within 30 seconds after
is
SPO2 Main
key.
MONITOR
Programming
% SpO 2
PULSE (BPM)
CHANNEL
SELECT
MONITOR
CHANNEL
OFF
Programming