Prepare Syringe and Administration Set (Syringe Module) (Continued)
CAUTION
Use compatible components which have the smallest
internal volume or "deadspace" to minimize residual
volumes between the syringe and the patient when
administering medications or fluids at low infusion rates
(e.g.,< 5 mL/h, and especially flow rates < 0.5 mL/h). This
reduces the amount of time it takes for fluid to reach the
patient, maintains delivery accuracy, and reduces occlusion
detection times
Ensure that the device is as close to level of patient 's heart
as possible. Patient's heart level should be in line with the
CHANNEL SELECT
Use the
PRIME SET WITH SYRINGE
Alaris Syringe module to speed up the engagement of the
module's mechanical components and decrease the
syringe's internal friction (see "Prime - Using Options
Menu").
If using a pre-run infusion practice (to allow for medication
equilibration prior to connection to the patient), ensure that
the distal end of the administration set is level with or higher
than the device.
Avoid use of manifolds with ports containing high pressure
valves. High pressure valves require additional pressure
(e.g., 50 - 200 mmHg) to open and allow fluid flow. These
high pressure valves may cause a significant delay in
therapy followed by a sudden bolus once the valve is
opened, particularly at low infusion rates (e.g., < 5 mL/h
and especially flow rates < 0.5mL/h).
These recommendations are especially important when
infusing high-risk or life-sustaining medications at low
infusion rates (for example, < 5 mL/h and especially at flow
rates < 0.5 mL/h).
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
key.
channel option on the
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART