The displays illustrated throughout this document are for
illustration purposes only. The display content varies,
depending on configuration settings, type of administration
set in use, hospital-defined Data Set uploaded using
the Guardrails Suite MX, programmed drug calculation
parameters, and many other variables. Refer to specific
drug product labeling for information concerning appropriate
administration techniques and dosages.
Main Display
See the PC unit section of this User Manual.
Dynamic Pressure Display
Dynamic Pressure Display
Current operating pressure is
indicated by solid bar.
Patient-side occlusion pressure
set point is indicated by tick mark.
The tick mark moves when the
Pressure Limit changes.
General Information
2-146
Features and Displays (Continued)

Displays

Midtown Hospital
Adult ICU
VOLUME
INFUSED
VTBI = 250.0 mL
AUDIO
ADJUST
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
CAUTION
Although the dynamic pressure
display bars for the Syringe module
and Pump module both use the full
width of the screen for display, they
each represent different ranges.
The Pump module's range is 50 to
525 mmHg and the Syringe module's
range is 25 to 1000 mmHg.
Pump and Syringe Modules Section