Prepare Administration Set (Pump Module) (Continued)
1.
Close roller clamp.
2.
Open Pump module door.
Set's safety clamp fitment automatically closes to prevent
accidental free-flow.
3.
Remove set, as follows:
a. Gently pull tubing below Air-in-Line Detector forward
and out.
b. Lift upper fitment from upper fitment receptacle.
4.
If set is being removed to begin a gravity flow:
a. Depress the blue ridged release tab on upper side of
safety clamp device.
b. Slide the white slide clamp into blue fitment (open
position).
c. Adjust the flow rate using set's roller clamp.
1.
Prepare primary solution container in accordance with
manufacturer's User Manual.
2.
Open administration set package, remove set, and close
roller clamp. (Refer to set's User Manual.)
3.
Insert administration set spike into prepared fluid
container, following accepted hospital/facility procedure,
and hang container 20 inches above Pump module.
4.
Fill drip chamber to
Getting Started
2-6

Remove

Prime

full.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
White Slide Clamp
(shown in closed position)

Hide quick links: