Programming with Interoperability—Subsequent Primary Infusion (Pump
The following procedures assume that the Pump module has
already been programmed with an Initial Primary Infusion.
NOTE:
If an infusion is in Delay when a subsequent automated
programming request is sent and matching criteria:
•
Are met, the Delay is canceled and the subsequent
automated programming request infusion parameters are
pre-populated on the PC unit. To add a Delay manually,
refer to Delay Options on page 2-102.
•
Are not met, the automated programming request is
rejected and the infusion remains in Delay.
The generic workflow for pre-population of infusion parameters
is as follows. Contact your EMR/HIS vendor for detailed
instructions.
1.
Using the EMR/HIS system scanner used for barcode
medication administration:
•
Scan the Patient's ID Band
•
Scan the fluid/medication barcode label
•
Scan the barcode label on the Pump module
2.
To indicate that information has been noted and to
continue programming, press the CONFIRM soft key on
the PC unit.
If the hospital has defined a Clinical Advisory for the drug, a
message will appear.
NOTE:
If the infusion parameters are unable to be pre-populated on the
Alaris System, error messages will be displayed in the EMR/HIS.
Error messages will not be displayed on the PC unit. Consult the
EMR/HIS.
3.
Start the applicable infusion as described in the following
procedures:
•
Continuous Infusion
•
Fluid Infusion
•
Intermittent Infusion
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Module) (Continued)
Programming with Interoperability and Guardrails Suite MX
A
Alteplase
Stroke
Clinical Advisory:
This dosing is for Acute
Ischemic STROKE
>Press CONFIRM
CONFIRM
2-73