Compatible Syringes; Features And Displays; Features And Definitions - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

The PCA module is calibrated and labeled for use with the
following single-use disposable luer-lock syringes. Use only the
syringe size and type specified on the Main Display. The full list
of permitted syringe models is dependent on the PCA module's
software version.
Syringe variability can impact occlusion pressure sensing. The
variability can reduce the device's time to alarm and/or might
require that a higher alarm pressure limit be programmed.
Manufacturer
20 mL
B-D Plastipak
IMS Pump Jet
Monoject
Terumo
NOTE:
Prefilled Morphine Sulfate 1 mg/mL.

Features and Definitions

See the PC unit section of this User Manual for system features and definitions.
Feature
Auto Pressure Limit
Adjustment
Auto Syringe
Identification
Bolus Delivery Rate
Bolus Dose
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015

Compatible Syringes

30 mL
35 mL
x
x
x
x
x
x
x

Features and Displays

When a bolus is delivered, pressure alarm limits are temporarily raised to
maximum limit.
System automatically detects syringe size and narrows down syringe
selection list.
Rate at which PCA, bolus, and loading doses (boluses) are infused.
Allows an additional amount of medication to be programmed once PCA
infusion has begun. Current PCA infusion resumes following delivery of a
Bolus Dose.
60 mL
x
x
x
Definition
General Information
3-49

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents