CareFusion Alaris 8015 User Manual page 424

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Alaris System (Continued)
3.
Follow the cleaner manufacturer's instructions on the time
to leave it on the device surface. Then, remove the cleaner
using a soft cloth dampened with water.
DO NOT allow the cleaner to collect on the instrument.
Cleaning the IUI
4.
Apply 70% isopropyl alcohol directly to the dedicated IUI
cleaning brush. To prevent cross-contamination, do not dip
the brush into the IPA.
DO NOT use the same brush used on the case to clean the
IUI connectors. Doing so could inadvertently transfer the
cleaner or contaminants to the electrical contacts.
5.
Clean both IUIs with the dedicated IUI cleaning brush.
To avoid accidentally depositing fluid on the connectors,
DO NOT use any spray cleaners anywhere near the IUI
connectors.
NEVER ALLOW ANY CLEANER OTHER THAN 70% IPA
TO CONTACT THE IUI CONNECTORS.
A-30
Cleaning (Continued)
Alaris System User Manual – with v9.33 Model 8015
Appendix Section

Hide quick links:

Advertisement

Table of Contents
loading