Bolus Volume, Maximum after
Occlusion:
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Specifications (Continued)
Syringe Module
WARNING
Installing a pressure sensing disc after an infusion has started can
result in a bolus to the patient.
Without Pressure Sensing Disc:
Pressure Setting
Low
Medium
High
With Pressure Sensing Disc:
Pressure Setting
300 mmHg
500 mmHg
1000 mmHg
Alaris System has a back-off safety feature that, when enabled and a
pressure sensing disc is in use, is designed to reduce bolus volume on
occlusion release.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Bolus Volume (mL)
0.512
0.776
1.103
Bolus Volume (mL)
Back Off Disabled
0.462
0.575
0.839
Back Off Enabled
0.126
0.331
0.323