CareFusion Alaris 8015 User Manual page 366

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

used. In the event of humidity or condensate outside the
EtCO
module's operating specifications, the EtCO
2
present a "Remove Blocked Disposable" message.
Due to the relatively small sampling size needed for EtCO
readings, partial pressure does not affect the ability of the
EtCO
module to measure EtCO
2
rate can be achieved.
Microstream Disposables are single-use, disposables which
must be changed with each use. The manufacturer's sample
flow, 50 mL/min, does not affect the disposable's life; however,
humidity and specific patient conditions can shorten the
effective life of the disposables. Microstream Disposables are
rated for up to 24 hours and 72 hours use, depending on the
specific Microstream Disposable.
The EtCO
module provides readings in compliance with BTPS
2
(body temperature, pressure, saturation) standards.
no affect on accuracy due to cyclic pressure up to 10 kPa.
NOTE:
BTPS (body temperature, pressure, saturation assumed 37°C,
47 mmHg) calculations are made according to:
PCO
= FCO
x (Pb - 47)
2
2
Where:
FCO
is fractional concentration of CO
2
FCO
= % CO
2
Pb is ambient pressure.
PCO
is partial pressure of CO
2
General Information
5-30
Principle of Operation (Continued)
, as long as the 50 mL/min
2
/100.
2
at BTPS.
2
module will
2
2
There is
in dry gas and
2
Alaris System User Manual – with v9.33 Model 8015
EtCO
Module Section
2