Pump Module - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Feature
Accumulated Air
Air-in-Line Settings
(single bolus)
Auto-Restart Attempts
KVO (Keep Vein Open)
Max Rate
Max VTBI
Pressure Mode
Mode Selection
Lock Status
Max Occlusion Pressure
Default Starting Occlusion
Pressure
SEC to PRI Alert
Secondary (Dual Rate
Sequential Piggybacking)
General Information
2-150
Configurable Settings (Continued)

Pump Module

Default Setting
Enabled
75 mcL
0
1 mL/h
999 mL/h
9999 mL
Pump
Unlocked
525 mmHg
525 mmHg
Enabled
Disabled
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Options
Enabled - Disabled
50, 75, or 250 mcL
Anesthesia Mode only: 500 mcL
0 - 9 attempts
Anesthesia Mode only: 9 attempts
-
mL/h
0.1
20
-
mL/h in
mL/h increments;
0.1
99.9
0.1
-
mL/h in
mL/h increments
100
999
1
-
mL
0.1
9999
Pump, Selectable
Locked, Unlocked
50 - 525 mmHg in 25 mmHg increments (adjustable
only in Selectable Pressure Mode)
50 - 525 mmHg in 25 mmHg increments (configured
by Profile and adjustable only in Selectable Pressure
Mode)
Enabled - Disabled
Enabled - Disabled
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents