CareFusion Alaris 8015 User Manual page 224

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Syringe Module (Continued)
Rate Accuracy:
Shock Protection:
Time to Alarm, Maximum:
Volume to be Infused
Programming Increments:
Weight:
NOTE:
The minimum Volume To Be Infused (VTBI) for both LVP and
Syringe module (applicable to all syringe sizes) is 0.1mLs. This
applies to Continuous Infusions, Bolus Dose, Intermittent
Infusions, and IV Fluids.
General Information
2-158
Specifications (Continued)
±2% of full scale plunger travel (not including syringe variation)
WARNING
Syringe size and running force, variations of back pressure, or any
combination of these can affect rate accuracy. Factors that can
influence back pressure are: Administration set configuration, IV
solution viscosity, and IV solution temperature. Back pressure can
also be affected by type of catheter. See "Trumpet and Start-Up
Curves" for data on how these factors influence rate accuracy.
Type CF, Defibrillator Proof
Rate (mL/h)
1
5
Maximum Time to Alarm specifications are based on following standard
operating conditions:
Atmospheric Pressure:
Back Pressure:
Disposable Type:
No Pressure Disc:
With Pressure Disc:
Humidity:
Syringe Type:
Temperature:
Range (mL)
0.01 - 9.99
10 - 60
4.5 lbs
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pressure Limit
No Disc
High Setting
(1000 mmHg) Setting
120 minutes
30 minutes
645 - 795 mmHg
0 mmHg before producing occlusion
#30914
#30920
20 - 90%
BD 50/60 mL
68 ±4°F
Increments (mL)
0.01
0.1
Alaris System User Manual – with v9.33 Model 8015
With Disc Highest
105 minutes
30 minutes
Pump and Syringe Modules Section