Adjust Audio Volume - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

1.
Press
AUDIO ADJUST
2.
To change volume to desired level, press either Louder
or Softer soft key. To sample alarm loudness level, press
soft key.
Test
3.
To return to PC unit screen, press
After 30 seconds without a key press, Main Display
appears.
NOTE:
The Minimum Audio Volume defaults to level 1. Levels 1-5 may be
set per profile. The user is able to lower the device AUDIO
VOLUME to the minimum limit set. When the minimum AUDIO
VOLUME LEVEL has been reached, the Louder/Softer soft keys
are unavailable and the PC unit emits an illegal key press audio.
General Setup and Operation
1-10
Start-Up (Continued)

Adjust Audio Volume

soft key.
MAIN SCREEN
WARNING
The user should check that the
current alarm preset is appropriate
prior to use on each patient.
WARNING
A hazard can exist if different Alarm
Presets are used for the same or
similar equipment in a single area,
such as an Intensive Care Unit or a
Cardiac Operating Theatre.
CAUTION
Setting the audio volume to the
lowest level will lower all system
alarms, including secondary alarms
soft key.
such as End of Infusion.
Alaris System User Manual – with v9.33 Model 8015
Midtown Hospital
Adult ICU
VTBI = 250.0 mL
VOLUME
AUDIO
INFUSED
ADJUST
Audio Volume Adjust
Test
3
Softer
Louder
>Change Setting or
Cancel
MAIN
SCREEN
PC Unit Section