Safety Clamp Fitment (Pump Module) - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Administration Set/Syringe Information (Continued)

Safety Clamp Fitment (Pump Module)

The primary administration set's safety clamp fitment is
a unique clamping device, on the pumping segment, that
prevents inadvertent free-flow when the administration set is
removed from the instrument.
Safety Clamp Fitment in Open Position
When a new Alaris Pump module/GEMINI Infusion System
administration set is removed from the package, the safety
clamp fitment is in the open position (white slide clamp aligned
with blue fitment). In this open position, flow is allowed as
required for the priming process. The roller clamp is used to
control flow during the priming process.
Safety Clamp Fitment in Closed Position
When an Alaris Pump module/GEMINI Infusion System
administration set is removed from the Pump module, the
instrument automatically engages the safety clamp fitment in
the closed position (white slide clamp projects out from under
blue fitment). In this closed position, flow is prevented.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Blue Fitment
White Slide Clamp
(shown in open position)
Blue Fitment
White Slide Clamp
(shown in open position)
General Information
2-135

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents