Alarm Message
SENSOR OFF
SENSOR WARMING
SWITCHOVER
STATUS
SYRINGE
CALIBRATION
REQUIRED
SYRINGE DRIVER
HEAD ERROR
SYRINGE EMPTY
SYSTEM ERROR
A-22
Alarms and Alerts (continued)
Alarms (continued)
Device
Priority
SpO
,
Medium
2
EtCO
Informational
2
Syringe,
Informational
PCA,
Pump
Syringe,
Medium
PCA
Syringe,
High
PCA
Syringe,
High
PCA
PC unit
High
Alarms
Meaning
The sensor is not connected to
the patient while monitoring or
after approximately 30 seconds
of searching for patient.
The monitoring device is
preparing sensor mechanisms to
search for and monitor the
patient vitals.
The device is delivering
medication/fluid.
The module needs calibration
before use. CALIBRATE scrolls
in the Message Display. Other
modules currently infusing will
continue to operate.
A noninfusing module, with the
plunger grippers open, senses
that excessive pressure is being
applied downward on the Drive
Head. The word "OCCLUSION"
scrolls in Message Display.
The syringe VTBI has completed
and the syringe contains less
than 2% total disposable
volume.
A FATAL MALFUNCTION is
detected on the PC unit
Alaris System User Manual – with v9.33 Model 8015
Response
Check the patient. Attach the sensor to
the patient.
N/A
No user respone expected.
Replace the module with an
operational device as soon as possible.
Service by qualified personnel is
required. Press the CONFIRM soft key
to continue.
To silence the alarm and continue with
normal operation, press the CONFIRM
soft key.
Set up a new syringe infusion or press
the CHANNEL OFF key.
Replace the device immediately. The
settings unrestorable; record the
settings before powering down the
device.