Alarm Message
CHECK SENSOR -
MOTION
INTERFERENCE
CHECK SENSOR -
WEAK PULSE
CHECK SENSOR -
WEAK SIGNAL
CHECK SYRINGE
CHECKING LINE
CLEARING
DISPOSABLE
CLOSE AND LOCK
DOOR
CLOSE DOOR, then
RESTART CHANNEL
Appendix Section
Alaris System User Manual – with v9.33 Model 8015
Alarms and Alerts (continued)
Alarms (continued)
Device
Priority
SpO2
Medium
SpO2
Medium
SpO2
Medium
Syringe
High
PCA
Pump
Informational
EtCO
Informational
2
PCA
High
Pump
High
Alarms
Meaning
The patient's motion has
inhibited monitoring.
The SpO
sensor is receiving a
2
weak pulse signal due to low
perfusion.
The SpO
sensor is receiving a
2
low quality signal.
The lever, plunger, and/or barrel
clamp is not engaged properly.
A patient-side occlusion has
occurred; the Auto-Restart
feature monitors downstream
pressure to determine if infusion
can continue.
The monitoring device is
attempting to clear disposable of
occlusion.
The PCA door is unlocked/open
during an active infusion. The
infusion is stopped.
The door was opened and
closed during an infusion. The
infusion stops on affected
channel.
Response
Check the sensor. Move the sensor to
a site with less motion.
Check the sensor. Confirm the correct
sensor placement. Move the sensor to
a better perfused site.
Check the sensor. Confirm the correct
sensor placement. Move the sensor to
a better perfused site.
Inspect the Syringe/PCA module and
adjust the appropriate sensor as
needed.
Reselect the syringe, restore or
reprogram and start or restart the
infusion.
No user response expected.
Wait for purging to complete.
Lock the door and press the RESTART
key.
Close the door. Press RESTART, or
press the CHANNEL SELECT key and
then press the START soft key.
A-15