Advertisement

Pump Module (Continued)
Flow Rate Programming
Increments:
Fluid Ingress Protection:
Infusion of Air, Means to
Protect Patient from:
Infusion Pressure,
Maximum:
KVO (Keep Vein Open)
Rate:
KVO Selection Range:
Occlusion Alarm
Thresholds:
Operating Principle:
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Specifications (Continued)
Rate Range
(mL/h)
User Input Rates
0.1 - 9.99
10 - 99.9
100 - 999
IPX1, Drip Proof
Ultrasonic Air-in-Line Detection
Maximum single bolus size = selectable 50, 75, or 250 microliters nominal
(500 microliters in Anesthesia Mode)
654 mmHg (Maximum Occlusion Alarm Threshold plus tolerance)
Factory default setting is 1 mL/h if set rate is 1 mL/h or above; or set rate, if rate
is 0.9 mL/h or below.
KVO rate can be set in System Configuration from 0.1 - 20 mL/h in 0.1 mL/h
increments.
Pump Mode: 525 mmHg at rates ≥30 mL/h
Varying level based on rate and patient back–pressure at rates
<30 mL/h.
Selectable Mode: User selected, 50 - 525 mmHg in 25 mmHg increments.
Positive displacement
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Increments (mL/h)
Device Calculated Rates
0.1
1
0.01
0.1
1
General Information
2-153