Pump Module (Continued)
Rate Accuracy:
Shock Protection:
Time to Alarm, Maximum:
Volume to be Infused
Programming Increments:
Weight:
NOTE:
The minimum Volume To Be Infused (VTBI) for both LVP and
Syringe module (applicable to all syringe sizes) is 0.1mLs. This
applies to Continuous Infusions, Bolus Dose, Intermittent
Infusions, and IV Fluids.
General Information
2-154
Specifications (Continued)
Rate accuracy of Alaris System is ±5% at rates between 1 and 999 mL/h and
±5.5% at rates <1 mL/h, 95% of the time with 95% confidence, under conditions
listed below.
Infusion Rate Range:
Ambient Temperature:
Source Container Height: 20 inches above top of Pump module
Test Solution:
Distal Back pressure:
Needle:
Administration Set Model: 2210
WARNING
Variations of head height, back pressure, or any combination of these
can affect rate accuracy. Factors that can influence head height and
back pressure are: Administration set configuration,
and IV solution temperature. Back pressure can also be affected by type
of catheter. See "Trumpet and Start-Up Curves" for data on how these
factors influence rate accuracy.
Type
, Defibrillator Proof patient applied part.
CF
Pressure Limit (mmHg)
50
50
525
525
Range (mL)
0.1 - 9.99
10 - 999.9
1000 - 9999
2.6 lbs
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
0.1 - 999 mL/h
68 ±4°F (20 ±2°C)
Distilled Water
0 mmHg (0 kPa)
18 gauge
Rate (mL/h)
1
25
1
25
Increments (mL)
0.01
0.1
Alaris System User Manual – with v9.33 Model 8015
solution viscosity,
IV
Time to Alarm
5 minutes
5 seconds
5 minutes
2 minutes
Pump and Syringe Modules Section