Guardrails PCA Pause Protocol Feature (Continued)
Program an Infusion (Continued)
3.
Review Clinical Advisory.
To continue, press
To activate PCA Pause Protocol, attach and start an
EtCO
module and/or SpO
2
protocol. To continue, press
4.
Verify correct parameters and press
confirm.
Prompt appears.
5.
Press
CONFIRM
6.
Start applicable infusion, as described in following
procedures:
Set Up PCA Dose Only
Set Up Continuous Infusion Only
Set Up PCA Dose + Continuous Infusion
Set Loading Dose Only
Programming
3-38
soft key.
CONFIRM
module per facility
2
CONFIRM
NEXT
soft key.
soft key.
soft key to
Alaris System User Manual – with v9.33 Model 8015
A
Clinical Advisory:
Attach an SpO2 or EtCO2
module now.
The drug selected will
activate the PCA Pause
Protocol once a monitoring
device is operating.
>Press CONFIRM
EXIT
CONFIRM
A
PCA Pause Limits
should be reviewed.
Press PAUSE LIMITS to
review the current settings.
Press CONFIRM to continue
programming the PCA
infusion.
>Press CONFIRM
PAUSE
CONFIRM
LIMITS
PCA Module Section