Select Pressure Limit; View And Clear Volume Infused - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

1.
Press
CHANNEL SELECT
2.
Press
OPTIONS
3.
Press Pressure Limit soft key.
4.
To select a pressure limit, press appropriate soft key.
5.
Press
CONFIRM
NOTE:
Option to change pressure limit can be selected:
after drug is selected, and before infusion mode is selected
and infusion starts, or
after infusion starts.

View and Clear Volume Infused

1.
To view volume infused, press
from Main Display.
Total volume infused, and time and date volume infused
was last cleared, is displayed for each channel.
Programming
3-36
Infusion Modes (Continued)

Select Pressure Limit

key.
key.
soft key.
VOLUME INFUSED
soft key
Alaris System User Manual – with v9.33 Model 8015
A
Channel Options
2 of 2
Pressure Limit - High
Prime Set with Syringe
>Select an Option or
EXIT
PAGE
EXIT
UP
A
Pressure Limit Selection
Low
Med
High
CONFIRM
PCA Module Section

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents