Manual Programming—Primary Infusion (Continued)
Intermittent Infusion (Continued)
If a dose outside of Soft Limits has been entered and
verified as correct, Message Display also shows either
"LLL" for a low dose or "
If a Soft Limit is overridden, G icon is displayed. When
G soft key is pressed, all applicable out-of-range limits
are listed.
5.
VTBI entry:
Pump module:
FU
When VTBI is less than 10 mL, entry can be to two
RATE (mL/h)
CHANNEL
SELECT
decimal places (one-hundredth of a mL).
PAUSE
CHANNEL
OFF
RESTART
VTBI is pre-populated with diluent volume of
infusion. To change VTBI, press
enter new value.
Syringe module:
If ALL Mode is enabled for syringe configuration
in Data Set,
estimated available volume in syringe is displayed.
FU
RATE (mL/h)
SELECT
CHANNEL
OR
PAUSE
CHANNEL
OFF
RESTART
If ALL Mode is disabled for syringe configuration in
Data Set, estimated available volume in syringe is
displayed when
To enter or change a numeric
soft key and enter value.
VTBI
To deliver entire contents of syringe: Keep an
ALL VTBI
numeric
VTBI
Manual Programming with Guardrails Suite MX
2-42
" for a high dose.

is displayed in VTBI field and
ALL
soft key is pressed.
VTBI
VTBI
value, or press
soft key to change a
ALL
value to
.
ALL
soft key and
VTBI
value, press
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Guardrails Drug Setup
Methotrexate
PRIMARY INFUSION
_ _ _
RATE =
mL/h
Available:
ALL
VTBI
(49.2 mL)
_ _ _ _
DURATION
h
min
1200
2
DOSE =
mg/m
[Conc]: 136.8 mg/mL
>Select DURATION
RATE
SETUP
VOLUME
Pump and Syringe Modules Section

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents