Infusion Modes; Programming Parameters - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Program an Infusion (Continued)
Drug Amount
Diluent Volume
OR
Drug Amount
Diluent Volume
If a Soft Limit is overridden, G icon is displayed. When
G soft key is pressed, all applicable out-of-range limits
are listed.
10. Verify correct parameters and press
confirm.
11. Prime syringe using Prime feature, if desired.

Programming Parameters

The PCA module uses the following programming parameters,
depending on infusion mode selected. See "General
Information," "Features and Definitions" for infusion mode
definitions and features.
PCA Dose: patient self-administered dose.
Lockout Interval: programmed time elapse between
availability of PCA doses.
Continuous Dose: basal rate dose.
Max Limit: (optional) total amount of drug which can be
infused over a specified time period.
Loading Dose: (optional) Bolus Dose infused prior to
initiation of PCA infusion.
Bolus Dose: (optional) additional dose programmed after
initiation of PCA infusion.
When the PC unit is in the Infusion Mode Selection, Infusion
Setup or Bolus Setup screens, a patient dose request from the
Dose Request Cord is handled as an unmet demand.
Programming
3-12
Prepare Infusion (Continued)
mg
1
mL
1
mg
30
mL
30
NEXT

Infusion Modes

soft key to
Alaris System User Manual – with v9.33 Model 8015
PCA Module Section