Features and Definitions (Continued)
Feature
All Mode
Auto Pressure
Auto Pressure Limit
Adjustment
Auto Syringe Size
Identification
Back Off
Fast Start
Infusion Complete
KVO
Near End of Infusion
(NEOI)
NOTE:
If the duration of the programmed infusion is less than or equal
to the configured limit set in the Guardrail Editor, the NEOI
alarm will occur at the start of the infusion.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Features and Displays (Continued)
When
is selected as the volume to be infused (VTBI), the entire
ALL
contents of syringe is delivered.
When enabled and a pressure sensing disc is in use, Auto Pressure option
is displayed in Pressure Limit screen. Auto Pressure automatically sets
alarm limit for a shorter time to alarm, as follows:
• If current pressure is 100 mmHg or less, system adds 30 mmHg to
current pressure to create a new alarm limit.
• If current pressure is greater than 100 mmHg, system adds 30% to
current pressure to create a new alarm limit.
When a bolus is delivered, pressure alarm limits are temporarily raised to
maximum limit.
System automatically detects syringe size and narrows down syringe
selection list.
This feature is only available when administration set in use has a pressure
sensing disc. When enabled, motor reverses plunger movement during
an occlusion until pressure returns to preocclusion levels, automatically
reducing bolus flow.
When Fast Start is enabled and an administration set having a pressure
sensing disc is used, instrument runs at an increased rate when an
infusion is first started, taking up any slack in drive mechanism.
An alert is given when current infusion is complete and VTBI has reached
zero.
If enabled, allows continuous infusions to automatically switch into KVO
mode upon completion. KVO option setting cannot be changed after
instrument is powered on and a Profile selected.
Allows an alarm to be configured to sound anywhere from 1 to 60 minutes
prior to the end of the infusion. The NEOI alarm can be configured
separately between continuous type infusions and intermittent infusions.
A separate alert time can be set for continuous and intermittent infusions.
Definition—Syringe Module
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
General Information
2-141