Pause Infusion (Continued)
3.
Press
soft key.
CONFIRM
scrolls in Message Display.
PAUSE
appears on Main Display.
PAUSED
Yellow Standby Status Indicator illuminates.
After 2 minutes:
and audio prompts begin, and yellow Standby Status
Indicator flashes.
4.
To reinitiate infusion:
Press
key.
RESTART
OR
Press
CHANNEL SELECT
NOTE:
The time displayed in the upper right corner of the screen is the
time of day in a 24-hour clock format (military time).
Since, by definition, a multidose infusion does not infuse for a
programmed period of time, it is assumed that another infusing
IV line keeps the vein open until the beginning of the first dose
and between subsequent doses. There is no keep vein open
(KVO) infusion at the completion of a programmed Delay until
infusion.
Syringe module: ALL Mode is not supported in Multidose
Mode.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
General Programming
2-108
Delay Options (Continued)
PAUSE - RESTART CHANNEL
key and then
START

Multidose Mode

visual
soft key.
WARNING
• The Multidose feature is to be
used only by personnel properly
trained in using multidose
infusions.
• Caution labels, which clearly
differentiate single dose and
multidose containers, must be
utilized.
• Single dose piggybacking
systems employing check
valve sets are not designed
for use with multidose
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents