Specifications and Symbols (Continued)
See the PC unit section of this User Manual for system symbols.
Symbol
Silenced alarm. Note: When displayed, the dashed lines through the symbol might
appear as solid lines due to the small display size.
Is displayed in Trend Data screen to identify an exceeded alarm limit.
Type BF patient applied part.
If the accuracy of any measurement does not seem
reasonable, first check the patient's vital signs by alternate
means and then check the SpO
functioning properly.
NOTE:
Because the SpO
2
distributed, only about two-thirds of the SpO
measurements can be expected to fall within ±Arms of the value
measured by a co-oximeter.
An inaccurate measurement can be caused by:
Incorrect sensor application or use.
Significant levels of dysfunctional hemoglobins (such as
carboxyhemoglobin or methemoglobin).
Intravascular dyes (such as indocyanine green or
methylene blue).
Exposure to excessive illumination; such as, a surgical
lamp (especially one with a xenon light source), bilirubin
lamp, fluorescent light, infrared heating lamp, or direct
sunlight.
Prolonged and/or excessive patient movement.
Venous pulsations.
General Information
4-28

Symbols

Measurement Accuracy

module to ensure that it is
2
module's measurements are statistically
Meaning
module
2
Alaris System User Manual – with v9.33 Model 8015
SpO
Module Section