Manual Programming—NO Guardrails Suite MX Protection (Continued)
Basic Infusion with Drug Calculation (Continued)
Pump module:
FU
When
VTBI
RATE (mL/h)
CHANNEL
SELECT
decimal places (one-hundredth of a mL).
PAUSE
CHANNEL
OFF
RESTART
In Drug Calculation mode, system infuses at
calculated rate rounded to nearest one-hundredth
of a mL per hour (as displayed on programming
screen). Rate shown in Rate Display is rounded to
nearest one-tenth of a mL per hour.
Syringe module:
If ALL Mode is enabled for syringe configuration
in Data Set,
estimated available volume in syringe is displayed.
FU
RATE (mL/h)
CHANNEL
SELECT
OR
PAUSE
CHANNEL
OFF
RESTART
If ALL Mode is disabled for syringe configuration in
Data Set, estimated available volume in syringe is
displayed when
To enter or change a numeric VTBI value, press
soft key and use numeric data entry keys.
VTBI
To deliver entire contents of syringe: Keep an
ALL VTBI
numeric
VTBI
soft key appears only if Bolus Dose is enabled
BOLUS
within selected Profile, drug is bolusable, and a VTBI is
entered.
15. Verify correct parameters and press
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
is less than 10 mL, entry can be to two
is displayed in
ALL
soft key is pressed.
VTBI
value, or press
soft key to change a
ALL
value to
.
ALL
START
field and
VTBI
soft key.
Programming—NO Guardrails Suite MX Protection
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A A A
Drug Calculation
CONTINUOUS INFUSION
4.2
RATE
mL/h
50
mL
VTBI
5
DOSE
mcg/kg/min
[Conc]: 5000 mcg/mL
>Press START
SETUP BOLUS START
PAUSE
2-91