Programming with Interoperability—Initial Primary Infusion (Continued)
The following workflow assumes that the Patient's ID band,
medication/fluid barcode label, and the module barcode labels
have been scanned, see Programming with Interoperability—
Initial Primary Infusion on page 2-62.
1.
Review and verify that all parameters pre-populated on the
Alaris System are correct prior to starting the infusion.
The rate and VTBI are pre-populated from the automated
programming request.
2.
Press the
START
NOTE:
Syringe module:
ALL Mode can still be enabled with interoperability.
However, the VTBI will pre-populate with the volume to be
infused from the automated programming request.
If the VTBI in the automated programming request is larger
than the actual volume in the syringe (for example, priming
the IV disposable with the syringe would decrease the
available volume in the syringe), the Alaris System will
display the VTBI field as blank and the message in the
prompt line on the Alaris PC unit would state: Value x
Exceeds MAX VTBI y. The clinician can either manually
program a value for the VTBI or use the ALL button if
appropriate (and if the ALL Mode is enabled in the data
set).
3.
For the Syringe module, unclamp the tubing and attach the
administration set to patient.
NOTE:
Unclamping tubing and starting infusion before attaching
administration set to patient minimizes any potential bolus that can
be released from pressure built up in set due to normal syringe
loading and priming.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Fluid Infusion
soft key.
Programming with Interoperability and Guardrails Suite MX
A
0.45% Sodium Chloride
PRIMARY INFUSION
100
RATE
mL/h
_1000
VTBI
mL
>Press START
DELAY
VOLUME
SECOND-
START
OPTIONS
DURATION
ARY
2-69