Manual Programming—Primary Infusion (Continued)
Intermittent Infusion (Continued)
6.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
•
To enter duration, press
numeric data entry keys (rate value is calculated and
displayed).
•
To enter rate, press
infusion rate.
7.
Verify correct parameters and press
•
If the programmed duration is outside the Soft Limit for
that care area, an audio alert sounds and a visual
prompt appears before programming can continue. If
Yes soft key is pressed, programming continues; if No
soft key is pressed, infusion needs to be
reprogrammed.
•
If the programmed duration is outside the Hard Limit
for that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
Infusion needs to be reprogrammed.
•
If a Soft Limit is overridden, G icon is displayed. When
G soft key is pressed, all applicable out-of-range limits
are listed.
•
For an Intermittent Infusion, the drug name and dose
will scroll on the module Message Display.
NOTE:
For an Intermittent Infusion, depending on how the drug library
is programmed, the dose may display as a total dose, drug
amount/kg or drug amount/BSA.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
soft key and use
DURATION
soft key and enter
RATE VOLUME
soft key.
START
A
Guardrails Drug Setup
Methotrexate
PRIMARY INFUSION
RATE =
VTBI
DURATION
1200
DOSE =
[Conc]: 136.8 mg/mL
>Press START
OR
DELAY
OPTIONS
VOLUME
Manual Programming with Guardrails Suite MX
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
4
mL/h
Available:
ALL
(49.2 mL)
6:00
hh:mm
2
mg/m
RATE
SETUP
START
2-43