CareFusion Alaris 8015 User Manual page 109

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Manual Programming—Primary Infusion (Continued)
Intermittent Infusion (Continued)
6.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
To enter duration, press
numeric data entry keys (rate value is calculated and
displayed).
To enter rate, press
infusion rate.
7.
Verify correct parameters and press
If the programmed duration is outside the Soft Limit for
that care area, an audio alert sounds and a visual
prompt appears before programming can continue. If
Yes soft key is pressed, programming continues; if No
soft key is pressed, infusion needs to be
reprogrammed.
If the programmed duration is outside the Hard Limit
for that care area, an audio alert sounds and a visual
prompt appears before programming can continue.
Infusion needs to be reprogrammed.
If a Soft Limit is overridden, G icon is displayed. When
G soft key is pressed, all applicable out-of-range limits
are listed.
For an Intermittent Infusion, the drug name and dose
will scroll on the module Message Display.
NOTE:
For an Intermittent Infusion, depending on how the drug library
is programmed, the dose may display as a total dose, drug
amount/kg or drug amount/BSA.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
soft key and use
DURATION
soft key and enter
RATE VOLUME
soft key.
START
A
Guardrails Drug Setup
Methotrexate
PRIMARY INFUSION
RATE =
VTBI
DURATION
1200
DOSE =
[Conc]: 136.8 mg/mL
>Press START
OR
DELAY
OPTIONS
VOLUME
Manual Programming with Guardrails Suite MX
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
4
mL/h
Available:
ALL
(49.2 mL)
6:00
hh:mm
2
mg/m
RATE
SETUP
START
2-43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents

Save PDF