Syringe Module - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

Feature
ALL Mode
Auto Pressure
Back Off (after occlusion)
Fast Start
KVO (Keep Vein Open)
Rate Adjust
Volume Adjust
Max Rate
Near End (NEOI)
Continuous Alert Time
Intermittent NEOI
Intermittent Alert Time
NEOI Snooze (Shared
Syringe and PCA setting)
Occlusion Pressure Set Point
With Disc
No Disc
Priming
NOTE:
*Occlusion Pressure Point cannot be set higher than the default
setting.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Configurable Settings (Continued)

Syringe Module

Default Setting
Disabled
Disabled
Enabled
Enabled
Disabled
1 mL/h
5%
999 mL/h
Disabled
60
Disabled
15
Disabled
1000 mmHg
High*
Disabled
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Enabled - Disabled
Enabled - Disabled
Enabled - Disabled
Enabled - Disabled
Enabled - Disabled
0.01 - 2.5 mL/h (0.01 - 0.09 mL/h available for 1 mL
and 3 mL syringes)
0.5 - 5% (Based on the total syringe size, not the
volume of fluid in the syringe)
0.1 - 99.9 mL/h in 0.1 mL/h increments;
100 - 999 mL/h in 1 mL/h increments
Enabled - Disabled
1 - 60 minutes
Enabled - Disabled
1 - 60 minutes
Enabled - Disabled
5, 10, or 15 minutes
-
mmHg in
mmHg increments
25
1000
1
Low (200 mmHg)
Medium (500 mmHg)
High (800 mmHg)
Enabled - Disabled
Options
General Information
2-151