Operating Features, Controls, Indicators (Continued)
IUI Connector, Left
Rate Display
Message Display
Channel (module) Identification
Channel (module) Select Key:
When pressed, selects corresponding
module for infusion parameter entry and
infusion setup.
Pause Key:
When pressed during
an infusion, temporarily stops infusion
on that module. After approximately
2 minutes, a visual and audio prompt
begins.
Channel (module) Off Key:
pressed and held until a beep is heard,
stops infusion on that module, deselects
that module, and if only that module had
been operating, system powers down.
Repeat for other operating modules to
power off each module.
Restart Key:
When pressed, resumes
operation of a previously paused or
alarmed infusion on that module.
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Features and Displays (Continued)
Status Indicators
Alarm
Infusing
(red)
(green)
When
Module Release Latch:
allows module to be removed.
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Standby
(yellow)
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pressure Transducer/Pressure Sensing Disc Housing
When pressed,
Gripper Control/Drive
Head Release
(shown in
closed position)
Plunger Grippers
(shown in closed position)
Barrel Flange Grippers
Syringe Barrel Sensor