Advertisement

Features and Definitions (Continued)
Feature
Lockout Interval
Max Dose Limit (Max
Accumulated Dose
Range)
Module Location
Enforcement
Near End of Infusion
(NEOI)
NEOI Snooze
Occlusion Pressure
PCA Module Section
Alaris System User Manual – with v9.33 Model 8015
Features and Displays (Continued)
A configurable interval of time (1-99 minutes) that must pass between PCA
Patient Doses. The lockout interval begins when the patient dose infusion
starts. The Lockout Interval range is 1-99 minutes.
Optional configuration that limits total amount of drug allowed to be
delivered to patient in a defined period (1, 2, or 4 hours).
• Should be configured in Data Set before Drug Library is developed.
Once drugs are in Profile PCA Drug Library, Max Accumulated Dose
Limit cannot be changed.
• Applies to all drug setups within Profile PCA Drug Library.
• Includes PCA Dose, Continuous Dose, and optional Bolus Dose if
selected to be included. Loading Dose is not included.
Tamper resistant security feature that ensures PCA Module is in a tamper
evident position. When enabled, PCA Module must be located to direct
right of PC Unit to allow programming an infusion.
The NEOI option allows an audio alarm to be configured based on a
percentage of the syringe size. The alert time can be set to occur when
5-25% of the syringe size is remaining.
Optional capability to remind users NEOI has been silenced. This is shared
Syringe and PCA setting.
Downstream occlusion alarm threshold can be set to low, medium, or high.
Definition
General Information
3-51