Programming with Interoperability—Subsequent Primary Infusion (Pump
Intermittent Infusion (Continued)
If the subsequent automated programming request
contains an infusion parameter (rate or dose) that affects or
changes the current infusion rate, a message appears with
a recalculation of the rate notification.
5.
From the
Infusion Setup
Select
to accept the rate change and start the
YES
infusion.
OR
Select
to return the infusion its current infusion rate
NO
and change the VTBI to the volume to be infused
contained in the automated programming request.
6.
Review and verify that all parameters pre-populated on the
Alaris System are correct.
7.
Press the
START
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Module) (Continued)
screen:
soft key.
Programming with Interoperability and Guardrails Suite MX
A
0.45% Sodium Chloride
PRIMARY INFUSION
100
mL/h
RATE
_1000
VTBI
mL
>Press START
DELAY
VOLUME
SECOND-
START
OPTIONS
DURATION
ARY
2-79