Manual Programming—Primary Infusion (Continued)
Continuous Infusion (Continued)
5.
An optional hospital-defined and editable starting value for
Continuous Infusion dose might already be entered.
OR
To make a rate or dose entry, press applicable soft key,
or
and use numeric data entry keys (other
RATE
DOSE,
value is calculated and displayed).
6.
To enter volume to be infused, press
enter value.
Pump module:
FU
When
VTBI
RATE (mL/h)
CHANNEL
SELECT
PAUSE
decimal places (one-hundredth of a mL).
CHANNEL
OFF
RESTART
In Drug Calculation mode, system infuses at
calculated rate rounded to nearest one-hundredth
of a mL per hour (as displayed on the programming
screen on the PC unit). The rate shown in Rate
Display on the Pump module is rounded to nearest
one-tenth of a mL per hour.
soft key appears only if Bolus Dose is enabled
BOLUS
within selected Profile, drug is bolusable, and a VTBI is
entered.
Manual Programming with Guardrails Suite MX
2-30
VTBI
is less than
mL, entry can be to two
10
OR
soft key and
Alaris System User Manual – with v9.33 Model 8015
FU
FU
RATE (mL/h)
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A A A
Alteplase
Stroke
CONTINUOUS INFUSION
_ _ _
mL/h
RATE
_ _ _ _
VTBI
mL
_ _ _ _ _
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Select Rate or Dose
SETUP
BOLUS
A A A
Alteplase
Stroke
CONTINUOUS INFUSION
56.7
RATE
mL/h
_ _ _ _
VTBI
mL
0.81
_
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Select VTBI
SETUP
BOLUS
Pump and Syringe Modules Section