Configurable Settings - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

See the PC unit section of this User Manual for system

configurable settings.

If the configuration settings need to be changed from the
Factory default settings, refer to the applicable Technical
Service Manual or contact CareFusion Technical Support
for technical, troubleshooting, and preventive maintenance
information.
With the Profiles feature enabled, the settings are configured
independently for each Profile. A hospital-defined, best-
practice Data Set must be uploaded to enable the Profiles
feature. Date and Time is a system setting and is the same in
all Profiles.
Models 8210
Feature
Limit Mode
Pulse Beep Volume
Pulse Rate Alarm Limit, High
Pulse Rate Alarm Limit, Low
SpO
Alarm Limit, High
2
SpO
Alarm Limit, Low
2
Feature
SatSeconds
SpO
Module Section
2
Alaris System User Manual – with v9.33 Model 8015
Configurable Settings
Default Setting
Adult Mode: 120 bpm
Neonatal Mode: 200 bpm
Adult Mode: 50 bpm
Neonatal Mode: 100 bpm
Neonatal: 95%
Adult: 90%
Neonatal: 80%
Model 8210
Default Setting
Off
and 8220
Adult
1
Adult: Off
Options
10, 25, 50, 100 seconds; Off
Options
Adult, Neonatal
1, 2, 3, Off
31 - 240 bpm
30 - 239 bpm
21 - 100%, Off
20 - 99%
General Information
4-23