Programming with Interoperability—Subsequent Primary Infusion (Pump
Continuous Infusion (Continued)
If the subsequent automated programming request
contains an infusion parameter (rate or dose) that changes
the current infusion rate, a message appears with a
recalculation of the rate notification.
4.
From the
Infusion Setup
Select
to accept the rate change and start the
YES
infusion.
OR
Select
to return the infusion to its current infusion
NO
rate and change the VTBI to the volume to be infused
contained in the automated programming request.
5.
Review and verify that all parameters pre-populated on the
Alaris System are correct prior to starting the infusion
6.
Press the
START
Pump and Syringe Modules Section
Alaris System User Manual – with v9.33 Model 8015
Module) (Continued)
screen:
soft key.
Programming with Interoperability and Guardrails Suite MX
A
Infusion Setup
New bag scan results
in a recalculation of the
rate to 45 mL/h
Accept rate change?
>Press Yes or No
Alteplase
A A A
Stroke
CONTINUOUS INFUSION
56.7
RATE
mL/h
100
VTBI
mL
DOSE
0.81
mg/kg/h
[Conc]: 1 mg/mL
>Press START
PAUSE
SETUP BOLUS
START
Yes
No
2-75