Displays; Configurable Settings - CareFusion Alaris 8015 User Manual

Alaris system with guardrails suite mx
Hide thumbs Also See for Alaris 8015:
Table of Contents

Advertisement

The displays illustrated throughout this document are for
illustration purposes only. The display content varies,
depending on configuration settings, type of disposable in
use, hospital-defined Data Set uploaded using the Guardrails
Suite MX, programmed parameters, and many other variables.
Main Display
See the PC unit section of this User Manual.
See the PC unit section of this User Manual for system

configurable settings.

If the configuration settings need to be changed from the
Factory default settings, refer to the applicable Technical
Service Manual or contact CareFusion Technical Support,
for technical, troubleshooting, and preventive maintenance
information. Please consult the facility's best practices
guidelines on the preferred alarm settings for the monitoring of
patients that are not continuously attended by a clinician.
With the Profiles feature enabled, the settings are configured
independently for each Profile. A hospital-defined, best-
practice Data Set must be uploaded to enable the Profiles
feature. Date and Time is a system setting and is the same in
all Profiles.
Al
Feature
t
Alarm Limit, High
E
CO 2
t
Alarm Limit, Low
E
CO 2
i
Alarm Limit, High
F
CO 2
Limit Mode
No Breath Alarm
EtCO
Module Section
2
Alaris System User Manual – with v9.33 Model 8015
Features and Displays (Continued)

Displays

Configurable Settings
Default Setting
Adult:
mmHg
60
Neonatal:
60
Adult:
mmHg
10
Neonatal:
10
Adult:
mmHg
8
Neonatal:
8
Adult
Adult:
seconds
30
Neonatal:
20
Options
-
5
99
mmHg
-
0
98
mmHg
-
2
99
mmHg
Adult or Neonatal
-
10
60
seconds
mmHg
mmHg
mmHg
seconds
General Information
5-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents